How to do a full reset in a UNIFI WorkGroup environment? - WKB17551
Article number: 17551
OBJECTIVE or GOAL
Reset a UNIFI server system.
ENVIRONMENT
- UNIFI network
PROCEDURE
- Log OFF all UNIFI portals.
- Switch OFF all Clients.
- Switch OFF LND.
- Switch ON the MS Pump Override.
- Switch OFF MS electronics.
- Switch OFF UPLC modules.
- Stop the w at servicehost service on the Server.
- Reboot the Server.
- Wait for all seven w at servicehost processes to start up on the Server.
- Switch ON LND.
- Wait five minutes for the three watservice host processes to start up on the LND.
- Switch ON UPLC modules.
- Switch ON the MS electronics.
- Wait five minutes for the MS to boot up.
- Switch ON the Client.
- Log in to the portal.
- Reset the LND in Device Manager.
- Check the vacuum in the instrument console.
- If the vacuum is good, switch the pump override switch back to AUTO. This should be reflected in the software.