Article number: 15855

SYMPTOMS

  • After creating and acquiring data with UNIFI, some blank injections are missing from the report

ENVIRONMENT

  • UNIFI 1.8.2 (workstation configuration)

CAUSE

Report is filtered to show only identified components.

FIX or WORKAROUND

  1. Change filters to show all components, or remove all filters so that all results are shown.
  2. Create a report based on the current view.

ADDITIONAL INFORMATION