Skip to main content
Waters

Cannot log in to UNIFI - WKB1372

Article number: 1372
This article is a portion of a Troubleshooting Path. To begin troubleshooting this issue, please select the link below.

SYMPTOMS

  • Cannot log in to UNIFI
  • Error message: Cannot connect to the UNIFI server because it is currently starting or unavailable

ENVIRONMENT

  • UNIFI 1.8

CAUSE

Oracle FRA (Fast_Recovery_Area) is missing, but is found on an external or USB drive.

FIX or WORKAROUND

  1. Re-attach the missing drive.
  2. Shut down Windows to power-off the computer.
  3. Power-on the computer and log in.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.