For UNIFI Network, are date/time stamps taken from the server, the LND, or both? - WKB114200

Article number: 114200

ENVIRONMENT

  • UNIFI 1.9.4 Network

ANSWER

Date/time stamps are taken from either the server or the LND as detailed below. Note that UNIFI is currently supported only on a local area network (LAN) where server, LNDs, and clients are synchronized with a time server and are located in the same time zone. See the P58 Installation and Configuration Guide (715005482 Rev E).

Review/Investigate Injections

  • Acquisition start time (LND)

Analysis Information  (Report Group)

  • Item(Set) Name (Server)*
  • Modified Date (Server)
  • Sample Set Created date (LND)

Explorer view (Analysis)

  • Set(Item) Name (Server)*
  • Modified Date (Server)
  • Created Date (Server)

Explorer view (Sample Set)

  • Set(Item) Name (Server)*
  • Modified Date (LND)
  • Created Date (LND)

*Dates are added to the automatic name generated for the analysis.

 