Cannot go into operate in UNIFI - MS error - microswitch error. - WKB10930

Article number: 10930

SYMPTOMS

  • Cannot put the MS into Operate
  • MS error for status
  • On the Diagnostics for the Lens Control PCB, there is a red LED for Interlocks indicating a microswitch error
  • If the MS is connected to MassLynx, no error exists and the instrument can go into Operate

UNIFI interlock error.PNG

ENVIRONMENT

  • UNIFI
  • Mass Spectrometer control

CAUSE

LND is locked up (even if it is a workstation)

FIX or WORKAROUND

  1. Go to Device Management and reset the workstation or LND if it is a work group environment. 
  2. Put the instrument into Operate. 