Alternative IP addresses are not assigned in Waters DHCP server configuration - WKB88721

Article number: 88721

SYMPTOMS

  • Customer Domain Network is running at 192.168.15.XX
  • Alternative IP addresses for LC and MS should be used to avoid conflicts
  • IP address of Instrument LAN card manually changed to 172.16.0.1 before software (MassLynx/UNIFI/Empower) was installed
  • Modules initialized without errors
  • No modules are visible in Waters DHCP Server configuration

ENVIRONMENT

  • UNIFI
  • Empower
  • MassLynx

CAUSE

  • IP address of instrument LAN Card was changed by hand before software was installed
  • Configuration guide for LAN card was not done inside the Waters DHCP Configuration Server

FIX or WORKAROUND

  1. Open Waters DHCP Server Configuration
    WDHCP alternative_1.JPG
  2. Go to Server-> "Configuration Wizard..."
    WDHCP alternative_2.JPG
  3. On the "Select Network Connection" page, click on the appropriate network connection for the instrument and then click "Next"

clipboard_e3bfec139cffd16400b7fe00704583530.png

  1. Change the IP address of the Instrument LAN to the alternate.
    WDHCP alternative_3.JPG
    WDHCP alternative_4.JPG
  2. Reboot the PC
  3. When the reboot finishes, start the modules

ADDITIONAL INFORMATION

https://support.waters.com/KB_Inf/MassLynx/WKB56069_Intermittent_IP_conflict_error

How to reset the 'hosts' file for communication to the Mass Spectrometer - wkb51278