Unable to process data in DynamX - WKB53369
Article number: 53369
SYMPTOMS
- Error appears when trying to process data: Error: Calculate MSResolution: Lock Mass Function absent or empty
ENVIRONMENT
- DynamX HDX Data Analysis Software 3.0
CAUSE
Data was acquired with the wrong Lock Mass acquisition setting (the "Acquire Lockspray - apply calibration" option is selected).
FIX or WORKAROUND
Re-acquire data, ensuring that the data is in continuum mode and that the "Acquire Lockspray - Do Not Apply Correction" option is selected.