With Agilent ICF 3.0, Column Compensation changes after saving the instrument method for a 7890 with FID - WKB46823
Article number: 46823
SYMPTOMS
- The Column Compensation for the FID changes to Column Compensation Curve #2 after saving the instrument method with "Nothing" selected
- The Column Compensation for the FID changes to Column Compensation Curve #1 after saving the instrument method with Compensation Curve #2 selected
- Same issues duplicated with in-house system

ENVIRONMENT
- Empower 3 Feature Release 4
- Agilent Instrument Control Framework (ICF) 3.0
- Agilent 7890 GC
CAUSE
Defect with A.02.04
FIX or WORKAROUND
- This does not happen with A.02.05. Upgrade to Agilent Instrument Control Framework (ICF) Update A.02.05.
or
2. Delete the default curve on the front panel of the 7890.
ADDITIONAL INFORMATION