MaldiChrom processing errors seen in HDI 1.4 - WKB46116
SYMPTOMS
- When processing data in HD Imaging (HDI) v1.4, errors are observed
- Example errors:
MaldiChrom v2.14 Feb-2016
Copyright 2008-2016 Micromass UK Ltd.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Tolerance' in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Flatten Edges’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Type’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(435)] Failed to find section [TOFACM MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini. [PROCPARS:gpps.cpp(445]] Failed to find key ’Centroiding Type’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process \MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445]] Failed to find key ’Use Areas’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
0 lockmass adjustments to be applied
MSE Level 0
ENVIRONMENT
- HDI v1.4 software
- SYNAPT G2-S/Si MALDI
CAUSE
These messages are normal and are a result of parameters not being specified in the .ini file. Default parameters are then used.
FIX or WORKAROUND
Do nothing. Messages are normal.