Error "Backup of the keystore failed" appears in the database backup logs in WDM - WKB252902
Article number: 252902
SYMPTOMS
- The following messages appear in the backup logs in WDM:
- copy Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\ewallet.p12 Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\ewallet.p12 /Y
copy Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\cwallet.sso Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\cwallet.sso /Y
Backup of the keystore failed
- copy Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\ewallet.p12 Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\ewallet.p12 /Y
- The backups are not completed
ENVIRONMENT
- WDM 4.x
- Oracle 19c
- NuGenesis
- Empower
CAUSE
The WATOracle or oraclejobs user does not have permission on the wallet files.
FIX or WORKAROUND
- Grant WATOracle or oraclejobs (as appropriate to the WDM version) full rights on the ewallet.p12 and cwallet.sso files.
- Grant the user full rights on the parent folder for the backupsets.
- Rerun the backup job (if using the run_it_now schedule) or wait for the next scheduled backup to complete.