LiveID PDF reports are generated incorrectly when compound names contain spaces - WKB197607
Article number: 197607
SYMPTOMS
- Empty pages where the top-to-tail plots should be
- The broken plots appear before plots that are visible
ENVIRONMENT
- LiveID 2.02303.604
CAUSE
The user created a custom library in which the names of compounds contain spaces, or created a report that should contain the top-to-tail plot of the one compound in the default library that contains a space.
FIX or WORKAROUND
Edit the library to remove spaces from all the names. For example, use "Phenylcyclidine_PCP" instead of "Phenylcyclidine PCP".
ADDITIONAL INFORMATION