Error: "[NG14510] Can't create temporary file" in files within the NuGenesis TE_TEMP folder - WKB8472
Article number: 8472
SYMPTOMS
- The following error message appears in NGD or mark_bad files in TE_TEMP:
- "[NG14510] Can't create temporary file. Please make sure that you have adequate disk space and appropriate permission to create files."
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
This issue may occur transiently during the normal operation of the NuGenesis File Capture modules or UNIFY. If this issue occurs repeatedly, after several RPC service restarts, then the NuGenesis RPC Service on the File Capture server does not have enough free disk space in TEMP.
FIX or WORKAROUND
- Restart the NuGenesis RPC service on the NuGenesis server and on the client
- If the issue persists, then follow the instructions in the linked article to move the RPC service's TEMP folder to a drive with more free space.