Error "Unknown Error" appears in the NuGenesis Archive Agent log file - WKB84615

Article number: 84615

SYMPTOMS

  • Empower projects are archived, but the JCAMP conversion step is skipped

ENVIRONMENT

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS
  • NuGenesis Data Adapters
  • Empower 3

CAUSE

The Empower project custom field "SampleName" was renamed to something else.

FIX or WORKAROUND

Rename the field back to "SampleName".

ADDITIONAL INFORMATION

The following messages appear in the Archive Agent log file Drive:\Program Files (x86)\Waters\NGSDMS\AA\NGSDMSAAActivity.log:

Message........: 2019-12-17 12:29:32: Activity message: Unknown Error:.
"E ","2019/12/17 12:29:39 PM ","-2147220989 ","-2147220989 ","Unknown Error: ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","NGPerformConvTasks ","1 ",""
Message........: 2019-12-17 12:29:32: EMC: PerformPreProcess error.
"M ","2019/12/17 12:29:53 PM ","0 ","0 ","Archived Project:[EMP3.analyse BNPG40 et BNVG40]  "," "," "," ","0 ",""