The message "Error during start of application.380:Invalid property value" appears when starting the NuGenesis SDMS archive template builder - WKB74945
Article number: 74945
SYMPTOMS
- The following error message appears when creating or editing an archive template:
- Error during start of application.380:Invalid property value
- After clicking OK to the error message, the main template builder window opens, but most of it is not functional
- No failures in the SDMS Administrator auto IQ log file
ENVIRONMENT
- NuGenesis 9 SDMS
CAUSE
The region formats in Windows are set to Portuguese (Brazil).
FIX or WORKAROUND
- Open the Windows Control Panel and double-click Region.
- Set the Format to English (United States).
- Click Apply, and click OK.
ADDITIONAL INFORMATION
Defect CRI-985 was filed for this issue.