Error "(121) - The semaphore timeout period has expired" appears in the NuGenesis Archive Agent activity logs - WKB73981
Article number: 73981
SYMPTOMS
- The following message appears in the NuGenesis Archive Agent activity logs:
- Message........: DATE TIME: GetFSInfofromOcx 10048 Err Desc - TRANSPORT ERROR GETTING VOLUME INFORMATION. Error Category: 1 System Code (121) - The semaphore timeout period has expired.
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
Semaphore timeout is caused by Server Message Block (SMBv1) being disabled on Windows 2016 Server.
SMBv1 is disabled with the standard install of Windows 2016 Server.
FIX or WORKAROUND
Enable SMBv1 on the Windows server.
ADDITIONAL INFORMATION
Follow the link below for a video that shows the steps to turn on Server Message Block (SMBv1).