Users are prompted for administrator credentials when previewing reports or restoring files from NuGenesis SDMS - WKB69283
Article number: 69283
SYMPTOMS
- Users are prompted for administrator credentials when opening reports or downloading files from NuGenesis.
- The application name as seen in the advanced tab is "NGipvw.exe".
ENVIRONMENT
- NuGenesis 9 SDMS
- Versions 9.0, 9.0.1
- User Account Control (UAC) is active in Windows
CAUSE
The NGipvw.exe program is incorrectly flagged in its internal manifest as requiring administrator privileges.
FIX or WORKAROUND
- The NuGenesis development team has supplied a version of NGipvw.exe that does not have this flag. Users are not prompted for administrator rights when using the modified NGipvw.exe. Contact Waters technical support for a copy of the hotfix
- Defect CRI-682 was filed for this issue
- This issue is fixed in NuGenesis 9.0.2 and 9.1
- The equivalent program in NuGenesis 8 - NG80ipvw.exe - does not have this flag in its manifest. NG8 is not affected by this issue.