Error message appears when releasing a request or selecting RequestReport in NuGenesis LMS - WKB57760
Article number: 57760
SYMPTOMS
- A long error message appears in the NuGenesis LMS client when releasing a request or selecting "RequestReport" from the Print menu. The relevant message text:
- original exception: ... Error evaluating expression for source text: $F{AMOUNT}+$F{UNIT}
- The report is not displayed
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
- The selected request's product category uses the default report template "RequestReport", either explicitly by selecting it in the category or implicitly by leaving the request report selection empty
CAUSE
A null value for the amount or unit metadata for one or more samples in the test request
FIX or WORKAROUND
- Defect report CRI-215 was filed for this issue.
- Restore the default template files for this report:
- Log in to NuGenesis LMS with read/write permissions on Reports
- Click LMS Main Menu > Administration
- Click the Reports button
- Select "RequestReport" in the list
- Click the button: Restore default report template and parameters
ADDITIONAL INFORMATION