SYMPTOMS

  • The following warning appears in the Export Utility's message log:
    • [NG14570] Unknown database error
  • The export utility does not export any records

ENVIRONMENT

  • NuGenesis 8 SDMS
  • SDMS 7.1
  • The project selected in the Export Utility is NOT a Managed Storage project

CAUSE

The search filter uses a wildcard expansion character: * or %.

FIX or WORKAROUND

  1. Redesign the search filter to avoid the use of wildcard characters. If the goal is to select all records in the project, configure a search filter like the following:
    • Record ID > 0

ADDITIONAL INFORMATION

 

id54684, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG