Error "502.3 - Bad Gateway" when loading NuGenesis WebVision - WKB3989
Article number: 3989
SYMPTOMS
- The following error message appears when loading WebVision:
- Server Error in Application [DEFAULT WEB SITE]
- HTTP Error 502.3 - Bad Gateway
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 9 SDMS
CAUSE
This issue occurs when an unsupported version of the Java Runtime Environment (JRE) is installed on a NuGenesis web server. The only version of Java supported for use with NuGenesis 8 is Java 1.7 Update 2 32-bit.
FIX or WORKAROUND
- Uninstall all JREs from the NuGenesis web servers, except for Java 1.7 Update 2 32-bit.
- Disable Java automatic updates on all NuGenesis web servers. For help, see the article How do I disable Java automatic updates on my NuGenesis web server?