Error "A connection must be established before attempting this operation" appears in the NuGenesis Archive Agent log when archiving Empower 3 projects - WKB32487
Article number: 32487
SYMPTOMS
- The following error appears in the Archive Agent log file for Empower-3 template lines:
- "A connection must be established before attempting this operation"
- No projects are archived from the Empower 3 systems
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters R3 or prior releases
- Empower 3 Feature Release 4 (FR4)
CAUSE
The Empower data adapters in NuGenesis 8 Data Adapters R3 and earlier versions are not compatible with Empower 3 FR4.
FIX or WORKAROUND
- Download NuGenesis 8 DA R3 Hotfix 1 from the Waters website and install it on the Empower 3 remote machine.
- The projects should be archived upon the next scheduled archive interval following the installation of the hotfix.