Error ORA-01461 occurs when saving some NuGenesis LMS documents or importing document templates - WKB310083
Article number: 310083
SYMPTOMS
- The following error message appears when saving some NuGenesis LMS documents:
- Original exception: java.sql.SQLException: ORA-01461: can bind a LONG value only for insert into a LONG column
ENVIRONMENT
- NuGenesis 9 LMS
- Chinese- or Japanese-language installation
CAUSE
Unknown
FIX or WORKAROUND
- Download the 21c ojdbc11.jar file from Oracle:
- Japanese: JDBC and UCP Downloads page | Oracle 日本
- Chinese: JDBC and UCP Downloads page | Oracle 中国
- Copy the ojdbc11.jar file to the LMS Server.
- Stop the "NuGenesis LMS Server" service.
- Rename Drive:\WatersLMSServer\Wildfly\modules\system\layers\base\waters\lms\oracle\main\ojdbc11.jar to ojdbc11.old.
- Copy the new ojdbc11.jar to the folder in step 4.
- Restart the "NuGenesis LMS Server" service.
ADDITIONAL INFORMATION
The new ojdbc11.jar will fail the IQ check in the LMS Server Auto IQ.