Files are missing from the Apache Tomcat and WebVision installation folders for a NuGenesis SDMS web server - WKB29240
Article number: 29240
SYMPTOMS
- The WebVision site is not accessible; the server returns an HTTP 502 error code and the web client displays an error message that the server proxy is not available
- On the NuGenesis server, the Apache Tomcat 6 service won't start, and it displays a "file not found" or "path not found" error message
- The Auto IQ log file for the NuGenesis SDMS Application Server (when run with admin rights) shows one or more missing files
- The "NuGenesis VISION Service" does not appear in the list of services
ENVIRONMENT
- NuGenesis 8 SDMS
CAUSE
Unknown
FIX or WORKAROUND
- Compare the files and folders in the Apache installation directory (Drive:\Program Files (x86)\apache-tomcat-6.0.29) with those on a NuGenesis server known to be good.
- Typically, the \bin\tomcat6.exe file and a few related DLLs will be missing—20 total files in the \bin folder instead of the expected 25 files—and the \lib folder will be missing.
- Run the NuGenesis SDMS Application Server Auto IQ with admin rights and look for missing files. Take note of the expected name and path of each missing file.
- Copy the files from the good server to the equivalent locations in the problematic server.
- Register the NuGenesis VISION service with the following command in a command prompt with admin rights:
- sc.exe create "NuGenesis VISION Server" binPath= "Drive:\Program Files (x86)\NuGenesis 8.0\Application Server\NG80wvls.exe" start= auto DisplayName= "NuGenesis VISION Server"
- Start the following services:
- Apache Tomcat 6
- NuGenesis VISION Server
ADDITIONAL INFORMATION