Error "Can't connect securely to this page" appears when loading NuGenesis SDMS WebVision - WKB237183
Article number: 237183
SYMPTOMS
- The following error message appears in the Web browser when loading SDMS WebVision:
- Can't connect securely to this page
- This might be because the site uses outdated or unsafe TLS security settings. If this keeps happening, try contacting the website's owner
- The WebVision site does not load, and the user has no option to bypass this message
ENVIRONMENT
- NuGenesis 9 SDMS
CAUSE
The Default Web Site in IIS on the SDMS Web Server is enabled for HTTPS, but it does not have a certificate selected for the site.
FIX or WORKAROUND
See article WKB17988: How to configure a NuGenesis Web server for HTTPS