Error "ORA-00942: table or view does not exist" when logging in to the NuGenesis Connectors Console - WKB221666
Article number: 221666
SYMPTOMS
- The following error message appears when logging in to the Connectors Console:
- [WSL 0107] The application cannot connect to the settings database. ORA-00942: table or view does not exist - Query: select PARAMETER, VALUE from CONNECTORS_SETTINGS..
- The Oracle error message may be translated from English depending on the NLS_LANG setting and/or the region and the language settings in Windows
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis Connectors 2.x
CAUSE
Connectors table not present in the database
FIX or WORKAROUND
Run the Database installer from the NuGenesis Connectors installation disc.
ADDITIONAL INFORMATION