How do objects in NuGenesis LMS map to objects in the NuGenesis Stability module? - WKB19051

Article number: 19051

ENVIRONMENT

  • NuGenesis 9 LMS
  • NuGenesis 8 LMS

ANSWER

NuGenesis Stability is the optional module for NuGenesis LMS that manages pharmaceutical drug stability studies. It uses the SLIM system from H&A Scientific to manage the stability studies. The NuGenesis Stability client has no means of creating objects such as user accounts, clients, products, sample templates, and other elements. These objects are created in NuGenesis LMS and uploaded to/synchronized with the Stability database. The following table maps the NuGenesis LMS objects and their labels with the equivalent objects in NuGenesis Stability:

NuGenesis LMS object Configuration point in LMS When uploaded/synchronized? NuGenesis Stability object
Organization LMS Administration > List > "SM_Organization" When list entry is saved Client
Formulation LMS Administration > Products After creation or edit of Product in LMS Formulation
Product LMS Administration > Products After creation or edit of Formulation in LMS Product
Test LMS Administration > Tests Definition After creation or edit of Test Definition in LMS Test categories
Results LMS Administration > Tests Definition After creation or edit of Test Result in LMS Test fields
Sample templates LMS Administration > Sample Templates When the Sample Template is approved in LMS Sample templates
Specifications LMS Administration > Specifications When the Specification is approved in LMS Properties and fields of Sample Templates
Locations LMS Administration > Locations After creation or edit of Location in LMS Labs
User accounts LMS Administration > User Management When user account is saved to LMS database Users
Methods LMS Administration > Methods When Sample Template or Specification which contains a test added from a method is approved "Default Method" metadata on Global Test Categories *

ADDITIONAL INFORMATION

* Methods are an LMS object that has no equivalent in Stability.  The method properties in LMS become metadata of the test categories in Stability after the tests are transferred.