"Invalid column number for alignment" error occurs when printing or previewing a TargetLynx report - WKB9616
Article number: 9616
SYMPTOMS
- "Invalid column number for alignment" error occurs when printing or previewing a report.
-
TargetLynx freezes. Printing is not possible and the application process has to be stopped in Task Manager.
-
The problem is seen randomly, in different sample batches.
ENVIRONMENT
- TargetLynx
- QuanLynx
CAUSE
This is a software bug - it's been around for years but it is only rarely reported and it's difficult to reproduce.
Development: We have looked into this and there is a problem with the underlying Microsoft grid control that we are using. There is no easy way to fix this besides re-writing our table control using a different grid control. Unfortunately, we don't have the resources to write a new table control.
FIX or WORKAROUND
Slightly change the widths of the columns to be printed
or
Change the printing orientation to Landscape.
or
Use a different QuanLynx Layout (.qlt) file