Cannot export MassLynx data to netCDF using Databridge in SYNAPT G2-S/Si or Xevo G2-XS SCNs - WKB9593
Article number: 9593
SYMPTOMS
- Databridge export to NetCDF fails and Databridge fails
ENVIRONMENT
- SYNAPT G2-S and MassLynx SCN 916 (and probably other SCNs for the SYNAPT G2-S)
- SYNAPT G2-Si and MassLynx SCN 957 (and probably other SCNs for the SYNAPT G2-Si)
- Xevo G2-XS QTof/Tof and MassLynx SCN 966, SCN 996, or SCN 1018
CAUSE
Data processing components of MassLynx, including Databridge, are not supported in many of the Tof instrument control SCNs.
FIX or WORKAROUND
Process the raw data using MassLynx desktop.
Databridge also works fine in instrument control SCNs for the Xevo G2 (e.g., SCN 990) and Xevo G2-S (e.g., SCN 971).
ADDITIONAL INFORMATION
Waters recommends that all data processing of Tof data, including use of Databridge, be carried out using the current desktop processing variant of MassLynx.