ELSD not communicating in MassLynx 4.2 - WKB91414
Article number: 91414
SYMPTOMS
Exhibits all or some of the following symptoms, depending on which ICS is loaded:
- DHCP Name and Type do not populate in DHCP
- Unable to ping the assigned IP address
- Firmware does not meet minimum requirement
- Unable to scan in the ELS Detector in the Inlet Configuration Wizard
- No firmware listed in the following folder path, C:\ProgramFiles(x86)\WatersInstruments\Firmware
ENVIRONMENT
- Windows 10
- MassLynx 4.2 SCN 971
- MassLynx 4.2 SCN 976
- MassLynx 4.2 SCN 985
- Auto Purification
- 2420 Evaporative Light Scattering Detector
- 2424 Evaporative Light Scattering Detector
- Acquity UPLC ELS Detector
- Xevo G2-S
CAUSE
Incompatible ICS
FIX or WORKAROUND
- Use Instrument Control Software (ICS) version 1.40 for the ACQUITY ELSD.
This can be found in Waters Driver Pack 4 Supplemental Release 1 (DP4SR1)
AcquityELSD 1.40.1882.1.
ADDITIONAL INFORMATION
id91414, 2424, 2424B, MLYNX, MLYNXV41, SCN971, SCN976, SCN985, SUPMM, UPELS, UPELSARC, XEVOG2SQTF, XEVOG2STOF