Skip to main content
Waters

How to configure a Agilent 1200-series LC in MassLynx - WKB8746

Article number: 8746

OBJECTIVE or GOAL

Configure MassLynx to control an Agilent 1200-series LC using Ethernet control.

ENVIRONMENT

  • MassLynx 4.1
  • MassLynx 4.2
  • Ethernet control

PROCEDURE

You need a contact closure cable to connect the contact closure PCB on the back of the Autosampler or pump (relay contacts) to the MS.

Ensure that the following events table is set up for every LC method:

 

Note that the software works only with A1200 firmware:

 

 

 

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.