Inlet method continues to run after the runtime in the method is achieved - WKB87182
Article number: 87182
SYMPTOMS
- Inlet method continues to run after the runtime is achieved
- Run time clock on the MasslLynx Status panel shows 0.00
- During the method runtime, Runtime status in the ACQUITY MS Console shows a time of 0 to runtime (e.g., 0 to 4 mins for a Runtime = 4 mins)
- After the runtime passes, runtime in the console shows a time range of runtime to a random time (e.g., 4 mins to 12 mins)
- In the inlet method, all parts of the method are set to 4 mins.
ENVIRONMENT
- MassLynx
- ACQUITY UPLC
CAUSE
One of the ACQUITY processes that run in the background is corrupted.
FIX or WORKAROUND
- Close MassLynx and the ACQUITY MS Console.
- Browse to C:\Program File (x86)\Waters Instrument\bin.
- Locate the StopMasslynxProcesses.cmd.
- Double-click on StopMasslynxProcesses.cmd to execute the command file.
- Open MassLynx and confirm communication with the system.
- Run sample to verify that the sample ends at the runtime specified in the method.
ADDITIONAL INFORMATION
id87182, comms, MLYNX, MLYNXV41, SUPMM, UPBINARY