No concentrations of analytes or internal standards are calculated in TargetLynx - WKB85611
Article number: 85611
SYMPTOMS
- No concentrations of analytes or internal standards are calculated in TargetLynx
- "Standards" can be found and have concentrations calculated
- The calibration curve is created
- Peaks for the compounds are present in the MassLynx chromatogram traces
- Peaks for the internal standard compounds are not present in the chromatogram traces
ENVIRONMENT
- MassLynx 4.2 SCN 966
- TargetLynx
CAUSE
Compounds are set to use Internal (relative - use internal standards), and there are no corresponding internal standard peaks found in the data traces.
Internal Standard compounds are set to External (absolute - no Internal Standard).
FIX or WORKAROUND
- Assess sample preparation, or determine why no internal standard should be present in those samples.
- Edit the processing method to use External (absolute - no internal standards) instead of Internal (relative - use internal standards).
- For the missing Internal Standard, the Analysis Response Type in Compound Properties must be set to Internal rather than External.