No concentrations of analytes or internal standards being calculated in TargetLynx - WKB85611
Article number: 85611
SYMPTOMS
- not able to get any concentrations to be calculated in the TargetLynx results file for "Analytes" or "Internal Standards" samples
- "standards" are able to be found and have concentrations calculated
- calibration curve is created
- peaks for the compounds are present in the MassLynx chromatogram traces
- peaks for the internal standard compounds are not present in the chromatogram traces
ENVIRONMENT
- MassLynx 4.2 SCN 966
- TargetLynx
CAUSE
Compounds are set to use Internal (relative - use internal standards) and there are no corresponding internal standard peak found in the data traces.
Internal Standards compounds are set to External (absolute - no Internal Standard)
FIX or WORKAROUND
- Check sample preparation or determine why no internal standard should be present in those samples or
- Edit the processing method to use External (absolute - no internal standards) instead of Internal (relative - use internal standards).
- For missing Internal Standard, Analysis Response Type in Compound Properties needs to be set to Internal rather than External.