No concentrations of analytes or internal standards are calculated in TargetLynx - WKB85611

Article number: 85611

SYMPTOMS

  • No concentrations of analytes or internal standards are calculated in TargetLynx
  • "Standards" can be found and have concentrations calculated
  • The calibration curve is created
  • Peaks for the compounds are present in the MassLynx chromatogram traces
  • Peaks for the internal standard compounds are not present in the chromatogram traces

ENVIRONMENT

  • MassLynx 4.2 SCN 966
  • TargetLynx

CAUSE

Compounds are set to use Internal (relative - use internal standards), and there are no corresponding internal standard peaks found in the data traces.

Internal Standard compounds are set to External (absolute - no Internal Standard).

FIX or WORKAROUND

  • Assess sample preparation, or determine why no internal standard should be present in those samples.
  • Edit the processing method to use External (absolute - no internal standards) instead of Internal (relative - use internal standards).
  • For the missing Internal Standard, the Analysis Response Type in Compound Properties must be set to Internal rather than External.