How do I inject from top to bottom of my plate in MassLynx? - WKB78706
Article number: 78706
OBJECTIVE or GOAL
Inject samples in vertical instead of horizontal sequence using MassLynx.
ENVIRONMENT
- MassLynx
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle
PROCEDURE
- Open the inlet method in MassLynx.
- Select Acquity Sampler > Plate Generator.
- Use the "Previous Rack" or "Next Rack" buttons to bring up the plate that will be used.
- Under Grid Reference, click the down arrow on the Referencing drop-down menu.
- Select Sequential Discontinuous (make sure the Horizontal First check box is cleared).
ADDITIONAL INFORMATION
The numbering format for the Bottle column on the MassLynx sample list will be Plate Number:Well Number. For example, to inject well/vial 7 on plate 1, input 1:7.
id78706, MLYNX, MLYNXV41, SUPMM, UPARBSMFTN, UPARHSMFTN, UPARPFTN, UPARSMFTN, UPLCCHIVD, UPSMPMGR