How to test the transform option in MassLynx - WKB77474
Article number: 77474
OBJECTIVE or GOAL
Test and verify that the Transform option is installed correctly in MassLynx.
ENVIRONMENT
- MassLynx
- Transform
PROCEDURE
- Open the default project in MassLynx.
- Open the Spectrum window.
- Select File > Open and load the file hfn1.raw.
- Click the cross in the toolbar to show the full spectrum of hfn1.raw.
- Verify that the Transform option is enabled in the Process menu.
- To test Transform, select Process > Mass Measure and use the default settings.
- MassLynx will create a centroid spectrum.
- Click on the centroid spectrum.
- Select Process > Find Manual.
- Click in the peak1 field, and then right-click on centroid peak 1081. The peak1 field should be populated with the correct mass.
- Click in the peak 2 field, and then right-click on centroid peak 1165. The peak2 field should be populated with the correct mass.
- Click the measure button. You should see mass 15126 reported in the window. Close this window.
- Click in the continuum spectrum and select Process > Transform.
- You should now see a new spectrum with mass 15127.
- If Transform is not enabled, the Transform option was not selected when MassLynx was installed.
ADDITIONAL INFORMATION
id77474, MLYNX, MLYNXV41, SUPMM