Skip to main content
Waters

How do I set up an ACQUITY UPLC sample manager to inject with 96-well plates in MassLynx? - WKB77253

Article number: 77253

OBJECTIVE or GOAL

Set up MassLynx to recognize 96-well plates for an ACQUITY sample manager.

ENVIRONMENT

  • MassLynx
  • ACQUITY UPLC Sample Manager 
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • ACQUITY I-Class Fixed Loop Injector
  • ACQUITY I-Class Flow Through Needle
  • ACQUITY H-Class Flow Through Needle

PROCEDURE

  1. Open the Inlet Editor in MassLynx.
  2. Under ACQUITY Sampler, select Bed Layout.
  3. Choose the Bed Layout to be used.
  4. Select the Edit Plates on the Layout tab.

clipboard_e45991705a95e2e93c57ebb0206e501eb.png

  1. Change the tray setting to the 96-well plate you wish to use and press OK (the top is tray 1 and the bottom is tray 2).

clipboard_e4da3e62784978f039ddf35abbb688eff.png

ADDITIONAL INFORMATION

 

id77253, MLYNX, MLYNXV41, SUPMM, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPLCCHIVD, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMPMGR

Not able to find a solution? Click here to request help.