Peak integration in TargetLynx results cuts off the bottom of the peak - WKB69746
Article number: 69746
SYMPTOMS
- Peak integration in TargetLynx results is incorrect
- Integration baseline is about one third the way up the peak
- Bottom part of the peak is cut off or excluded from the integration
ENVIRONMENT
- MassLynx
- TargetLynx
CAUSE
The retention time window setting in the TargetLynx method is too narrow.
FIX or WORKAROUND
- Measure the width of the peak at the base in the Chromatogram view.
- Open the chromatogram and display the specific MRM trace.
- Zoom in on the peak of interest.
- Hold down the right mouse button and drag across the base of the peak.
- Continue to hold the right mouse button and take note of the time shown in the lower-left corner of the Chromatogram view.
- Divide this number by 2 and add about 10%.
- Type this result into the Retention Time Window setting on the Compound Properties page of the TargetLynx method for this compound.
- Set the Locate Peak Selection setting to Largest.
- Reprocess the dataset.
ADDITIONAL INFORMATION