MassLynx does not inject sample; "Missing/wrong plate or vial" in IntelliStart logs; system runs analysis and keeps injecting samples - WKB69236
Article number: 69236
SYMPTOMS
- Vial is accidentally placed in the wrong position in the rack or is not put in at all
- MassLynx cannot inject the sample but runs the analysis, collecting data (empty run, chromatogram with no peaks), running inlet method gradient, and so on
- No message appears
- Error seen in IntelliStart logs: "Missing/wrong plate or vial"
ENVIRONMENT
- MassLynx
- ACQUITY Sample Manager
CAUSE
The Sample Manager Preference is set to continue after warning.
FIX or WORKAROUND
- If you want the system to stop on error, configure the Sample Manager to stop on vial error change.
- In the console, click Sample Manager > Configure > Preferences.
- For "When a vial is missing:" select "Injection will fail on error".