No chromatograms in TargetLynx - WKB68785

Article number: 68785

SYMPTOMS

  • No chromatograms are generated in TargetLynx when processing a MassLynx sample list

ENVIRONMENT

  • MassLynx
  • TargetLynx

CAUSE

Function numbers have changed in the MS method and are therefore no longer associated with the function numbers in the processing method.

FIX or WORKAROUND

  1. Open the TargetLynx method.
  2. Change all function numbers to zero.
  3. The processing method will now use the compound's quantification trace, instead of the function number, as the identifier.

TargetLynx.JPG

ADDITIONAL INFORMATION