"Collection control not configured" after reloading MassLynx - WKB6738
Article number: 6738
SYMPTOMS
- MassLynx reloaded
- "Collection control not configured" error appears
ENVIRONMENT
- MassLynx
- 2767 Injector/Collector
- Autopure system
CAUSE
The collection control file was not selected.
FIX or WORKAROUND
- Copy the FractionLynx Configuration File (.FL4) from the MassLynx backup folder to the new MassLynx folder.
- Open MassLynx.
- Click FractionLynx.
- Click Collection Control.
- Select File > Open.
- Browse to the FractionLynx Configuration File (.FL4) in the new MassLynx folder.
- Click Open.
ADDITIONAL INFORMATION
To create a new FractionLynx Configuration File, see the Help topic "Creating a Fraction System". Access Online Help by using the "Help" menu in MassLynx or by pressing F1.