Cannot unpause the queue in MassLynx after using the Emergency Collection in FractionLynx - WKB67128
Article number: 67128
SYMPTOMS
- After pressing the "Emergency Collection" function in FractionLynx, the queue stays paused in MassLynx.
- Pressing the Pause button does not unpause the queue.
ENVIRONMENT
- FractionLynx
- MassLynx
- Autopurification
CAUSE
This is normal behavior for FractionLynx.
FIX or WORKAROUND
- Open FractionLynx.
- Go back to the Emergency collection while in the System view under Run.
- Click the Resume button.
- You can now unpause the system.
ADDITIONAL INFORMATION