MassLynx starts acquiring without triggering the inlet to inject - WKB65077
Article number: 65077
SYMPTOMS
- When starting a sample list run in MassLynx, the MS starts acquiring immediately.
- Cannot start the injection sequence.
- The inlet does not inject the sample.
- Inlet Events and Triggering are configured correctly.
ENVIRONMENT
- MassLynx
- Any Mass Spectrometer with LC inlet
CAUSE
Inlets not configured correctly/wrong modules selected.
FIX or WORKAROUND
Reconfigure the inlet:
- De-configure the inlet in the inlet editor by clearing all the modules.
- Restart the PC.
- Open MassLynx, and then the inlet editor.
- Reconfigure the inlet system.
ADDITIONAL INFORMATION