MassLynx starts acquiring without triggering the inlet to inject - WKB65077

Article number: 65077

SYMPTOMS

  • When starting a sample list run in MassLynx, the MS starts acquiring immediately.
  • Cannot start the injection sequence.
  • The inlet does not inject the sample.
  • Inlet Events and Triggering are configured correctly.

ENVIRONMENT

  • MassLynx
  • Any Mass Spectrometer with LC inlet

CAUSE

Inlets not configured correctly/wrong modules selected.

FIX or WORKAROUND

Reconfigure the inlet:

  1. De-configure the inlet in the inlet editor by clearing all the modules.
  2. Restart the PC.
  3. Open MassLynx, and then the inlet editor.
  4. Reconfigure the inlet system.

ADDITIONAL INFORMATION

 

id65077, MLYNX, MLYNXV41, SUPMM