How do I view the ratio of the analyte area to the internal standard area in TargetLynx? - WKB6247
Article number: 6247
OBJECTIVE or GOAL
To view the ratio of the analyte area to the internal standard (IS) area in the TargetLynx browser.
ENVIRONMENT
- MassLynx
- TargetLynx
PROCEDURE
Use the peak response field.
peak response = analyte area / IS area x IS concentration
When the concentration of the internal standard (IS) is set to fixed 1.0:
peak response = analyte area / IS area
- Select the internal standard in the left pane of the TargetLynx method.
- Click Compound properties.
- Set the Response Type to External (absolute - no internal standards).
- Click Calibration properties.
- Set Propagate Calibration Parameters? to No.
- Set Concentration of Standard: level to Fixed.
- Set Concentration of Standard to 1.0.
- Select the compound in the left pane.
- Click Compound properties.
- Set Response Type to Internal (relative - use internal standards).
- Select the internal standard in the Internal Standard: 1 field.
- Save the method.
- Process samples.
- Right-click in the summary table, and then select Change Column Order.
- Highlight Peak Response in the left pane and click Add.
- Click OK to close. The response column appears in the summary table.
- Select File > Save Layout As to save the browser layout.
ADDITIONAL INFORMATION
