How to save a calibration curve in TargetLynx? - WKB6189
Article number: 6189
OBJECTIVE or GOAL
Save a calibration curve to use in processing other samples.
ENVIRONMENT
- MassLynx
- TargetLynx
PROCEDURE
- Process standards with the sample type Standard.
- In the TargetLynx browser, select File > Export > Calibration.
- Save the calibration curve file (*.cdb) in the CurveDB folder of the appropriate project.
ADDITIONAL INFORMATION
To process samples using the saved calibration curve, see How do I process samples using a saved calibration curve in TargetLynx?