MassLynx becomes unresponsive and requires EPC reboot with Xevo TQ-S micro - WKB61832
Article number: 61832
SYMPTOMS
- MassLynx stops, and no further data is acquired
- Must reboot EPC before the system works again
- MassLynx 4.2 SCN 977 and onward
ENVIRONMENT
- MassLynx
- Xevo TQ-S micro
CAUSE
Bug in MassLynx software: an issue with certain dwell times causes a timing conflict. Reported in CRI-6758.
FIX or WORKAROUND
- Fixed in SCN 1049
- Workaround: Do not use the following dwell times:
0.041 s, 0.169 s, 0.233 s, 0.338 s, 0.402 s, 0.658 s, and 0.858 s
ADDITIONAL INFORMATION
Dwell times from 1 ms to 1 s were tested; only the times noted above caused an issue.
This is the same as the described in
https://support.waters.com/KB_Inst/M...ts_acquisition