Pump flow stops almost immediately after inlet method is downloaded to the system - WKB57357
Article number: 57357
SYMPTOMS
- Pump flow stops almost immediately after inlet method is downloaded
- Changing method from Analytical mode to Prep mode
- Correcting method error—column position originally set to analytical is set to prep column—does not resolve issue
- Found: Loop selection set to analytical (left) in the Autosampler section of of the inlet method
ENVIRONMENT
- 2525/2545 Binary Gradient Module
- 2767 Sample Manager
- System Fluidics Organizer
- MassLynx 4.1
CAUSE
Loop selection is set to the wrong position in the Autosampler parameters for the inlet method for prep runs.
FIX or WORKAROUND
- Select the "Right" valve in the Loop selection section of the Autosampler parameters for the inlet method for prep runs.
- Save the updated inlet method.
- Download the inlet method to the system.
ADDITIONAL INFORMATION