ACQUITY Console does not open - WKB5649

Article number: 5649

SYMPTOMS

  • ACQUITY console does not open
  • Found: Rebooting ACQUITY hardware, MassLynx PC, and Quattro Micro EPC does not resolve issue
  • Found: Reloading MassLynx 4.1 and ACQUITY system drivers does not resolve

ENVIRONMENT

  • MassLynx 4.1 SCN 805
  • ACQUITY System June 2011 Driver Pack
  • Quattro Micro
  • ACQUITY Classic
  • Windows XP Service Pack 3

CAUSE

Incorrect network card and operating system configuration

FIX or WORKAROUND

  1. Reimage the PC with a clean installation of Windows.
  2. Configure Windows XP as recommended by Waters.
  3. Install MassLynx 4.1 SCN 805.
  4. Install ACQUITY system drivers for the ACQUITY Classic components.

ADDITIONAL INFORMATION