Calibration curve (.cdb) file is not applied to response in TargetLynx processing, resulting in no reported concentation values - WKB56408
Article number: 56408
SYMPTOMS
- No concentration values reported in TargetLynx results
- Peak areas are all reported correctly
- Peak responses are all calculated
- Calibration curve file is applied during processing
- Correct calibration file is shown at the top of the Calibration panel in the TargetLynx browser
- Found: data processed with Calibrate Standard enabled gives correct and complete results
- Found: reported symptoms are reproduced using a dataset in the Targetlynx.pro project
ENVIRONMENT
- MassLynx 4.2
- TargetLynx XS
CAUSE
Corruption in TargetLynx processing
FIX or WORKAROUND
- Close MassLynx, all of its windows, and the TargetLynx Browser.
- Go to START > Run and specify REGEDIT in the Run dialog.
- In the Registry Editor, expand the folder tree on the left-hand side, and then browse to the following two folders and delete them:
- HKey_Current_User > Software > Micromass >> Locate and delete the Targetlynx XS folder
- HKey_Local_Machine > Software > Wow6432Node > Micromass >> Locate and delete the Targetlynx folder
- Open MassLynx.
- Open a sample list in the project for the standard curve samples.
- Process the standard curve samples to create a new TargetLynx Dataset (QLD) file for the standard curve.
- Save the QLD file with an appropriate name. (You cannot export methods or results until the QLD file has a name.)
- Go to File > Export > Calibration. Save the calibration curve information in the Curvedb folder of the project.
- Process a set of samples applying the newly created calibration curve file.
Note: The TargetLynx and TargetLynx XS registry key folders will be recreated the first time TargetLynx runs.