Are the M-Class and nanoACQUITY supported with MassLynx and Windows 10? - WKB55605

Article number: 55605

ENVIRONMENT

  • MassLynx 4.2
  • Windows 10
  • nanoACQUITY UPLC
  • ACQUITY UPLC M-Class
  • Windows 10

ANSWER

ACQUITY M-Class is supported under Windows 10 using Waters Driver Pack 2019 Release 2 (DP 2019 R2) for Empower and MassLynx (Microscale). This is downloadable from Waters.com.

The nanoACQUITY was tested on Windows 10, and the release notes will be re-released shortly to reflect this.

Be sure to use the microSM [v1.56] firmware for the nanoACQUITY SM and the microSM-FL [v1.56] firmware for the M-Class SM.
Mclass firmware ICS loader 2.png

See the release notes for full details of Waters Driver Pack 2019 Release 2 (DP 2019 R2) for Empower and MassLynx (Microscale).

Find installation instructions here.

 