BiopharmaLynx error "Processor failed to process" - WKB54882
Article number: 54882
SYMPTOMS
- When trying to run a search with BiopharmaLynx, the error "Processor failed to process" appears
ENVIRONMENT
- BiopharmaLynx
- MassLynx
CAUSE
The lockmass function in the MS method was set to "acquire lockmass and correct", so data was lockmass corrected before being given to BiopharmaLynx. BiopharmaLynx prefers uncorrected data so that it can use lockmass data and its own algorithm to generate and apply lockmass correction.
FIX or WORKAROUND
Collect lockmass data for BiopharmaLynx processing, but make sure to NOT apply the lockmass correction while acquiring data.
ADDITIONAL INFORMATION