How do you extract and sum individual mass chromatograms from full scan data in MassLynx? - WKB5476
Article number: 5476
OBJECTIVE or GOAL
Extract and sum individual mass chromatograms from full scan data.
ENVIRONMENT
- MassLynx
PROCEDURE
- Open the data file in the Chromatogram window in MassLynx.
- Go to Display > Mass.
- In the Description (m/z) field, specify the following:
- For individual mass chromatograms, enter the m/z of interest (example: 300).
- To sum multiple mass chromatograms, enter the m/zs, separated by a plus (+) sign (example: 300+400+500).
ADDITIONAL INFORMATION
This format is applicable to the Quantification Trace and Target Ion Trace fields in QuanLynx and TargetLynx.