Skip to main content
Waters

Using MassLynx SCN 714 or SCN 805, no data acquired via "Analog In" - WKB45569

Article number: 45569

SYMPTOMS

  • It is not possible to acquire an analog signal using the "Analog-In" connection on the back of the mass spectrometer. No analog data files are acquired.
  • If an analog channel is enabled and the MS method saved and then reopened, the analog channel is disabled. Also, the word "Enabled" is added to the description field.
  • Every time the MS Method is saved, closed, and reopened, every channel has "Disabled" or "Enabled" added to the Description field for the channel, depending on whether the check box was selected or clear when the file was closed. The Disabled/Enabled text in the Description field grows in length, as shown below:

clipboard_ec16257ce63ab61a439459cf499e10196.png

 

ENVIRONMENT

  • MassLynx SCN 805
  • MassLynx SCN 714
  • Quattro micro
  • Quattro Premier
  • ZQ source

CAUSE

Software defect - will not be fixed

FIX or WORKAROUND

The workaround for the Quattro micro GC is to acquire via Ethernet using the phase 2 A7890 GC ICS.

The workaround for other instruments:

  1. Downgrade MassLynx 4.1 to SCN 627 (do not downgrade to SCN 714).
  2. Collect data via an eSAT/IN.
  3. Each time the method is edited, clear all of the Disabled/Enabled text from the description field before closing and saving the MS method.

ADDITIONAL INFORMATION

The eSAT/IN is not included in the SCN 805 Deployment Manager, so you must install the ICS manually as follows: 

Go to C:\Program Files (x86)\Micromass Utilities\ICOP\eSATIN and run the setup.exe file.

id45569, MLYNX, MLYNXV41, QUATPREMXE, QUMICRO, QUMICROGC, QUPREMIER, SCN627, SCN714, SCN805, SUPMM, ZQ2000, ZQ2000SFC, ZQ4000

Not able to find a solution? Click here to request help.